Chapter 5
56
ALARMS
The Perfusor® Space is equipped with a audible and optical alarm signal.
Audible
Alarm-
type
signal Red LED
Device
yes
flashes
Alarm
Pre-
yes
off
Alarm
Reminder
yes
off
Alarm
Alarm
no
off
Hint
6.1 Device Alarms
When a device alarm occurs the infusion is immediately stopped. Press
off the device. Then switch the device on again. In case of a repeated device alarm you
must disconnect from the patient, open the front door of the pump and take out the
disposable. The device needs to be handed to the service department.

6.2 Pre-Alarms and Operating Alarms

Pre-alarms:
Pre-alarms occur a few minutes (dependable on service settings, excluding OccluGuard
and pressure leap/drop pre-alarms) prior to operating alarms. During pre-alarms an au-
dible tone sounds, the yellow LED is constantly on and a staff call is activated (optio-
nal). The display message varies depending on the alarm reason. The signal tone and
the staff call are turned off with
rating alarm goes off. Pre-alarms don't lead to an interruption of the infusion.
Optical signal
Yellow LED Text
off
device
alarm and
alarm code
(see service
program)
constant
see alarm
on
description
constant
see alarm
on
description
off
see alarm
description
c
. Display and LED stay in pre-alarm until the ope-
Staff call
User confirmation
O
yes
Press
and follow
the instruction on the
display.
c
(de-)activate
Press
via service
alarm and turn off
program
staff call. Visible
alarm remains until
end.
k
yes
Press
to mute
alarm, turn off staff
call and delete the
alarm text.
no
Hint disappears
without confirmation.
O
ALARMS
to mute
to switch