Chapter 3
Note:
and Trumpet Curves.
Note:
(e.g. smallbore extension set) or use a back check valve for lines at the same
access which are not used for Take Over Mode.
Note:
will be turned on again, both will start as a normal pumps and not in TOM
modus with the latest TOM setting.
When both pumps are put in Standby during infusion and both are turned on
again, both pumps will start with the latest druglibrary setting. TOM has to be
activated separately.
TOM Requirements:
TOM will only be offered if the following requirements are met:
• Same drug selected on both pumps.
• Same drug concentration selected on both pumps.
• First Perfusor® Space pump must have a running infusion with a drug
TOM Hints:
The following TOM hints are to be observed:
Start-up behaviour is not influenced by TOM. See Chapter Start Up Graphs
Please use a seperate patient connection for Take Over Mode infusion
When both pumps are turned off (e.g. for changing the syringe), pumps
from drug library.
SPECIAL FUNCTIONS
43