Chapter 4
Note:
• Once all values are confirmed, the Main Menu is displayed.
Note:
library, soft limit may be overridden or value re-programmed per institutional
policy. Order will be rejected if hard limit is exceeded. (except in circumstance
where pump service program is not set to perform drug library match for auto-
programming).
For PRIMary Orders (Either 'Continuous' oder 'Dose over Time'):
Note:
PRIMary infusion, subsequent orders will be considered PIGGYback.
Note:
no subsequent order can be received. Additional, no updates can be received for
'Dose over Time'.
• Press Start/Stop key
Updates to Current Primary Infusion
Updates may be received for PRIMary infusions while pump is running or stopped and
while in PRIMary or PIGGYback.
While in PRIMary:
• Update icon will appear on display, follow on screen prompts to accept or cancel
the order. Confirmation screen will indicate both OLD and NEW value for
parameter(s) that changed.
While in PIGGYback:
• Message will appear on top of display indicating update is available for PRIMary.
• Press
Order may be cancelled prior to confirming order.
Soft Limit alert will be issued if value exceeds any soft limits set in drug
The first order sent send as 'Continuous' is always considered as the
Order sent as 'Dose over Time' is always considered the PRIMary infusion,
f
s
to start infusion.
M
key to view order.
AUTOPROGRAMMING
47