Chapter 6
"Calibrate device"
"Claw malfunction"
"Plunger plate not prop. fixed" The syringe plunger plate does not attach to the
"Standby Time expired"
"No battery inserted"
"Data were reset"
"Therapy data were reset"
"Data Lock"
"Connection lost – TOM aborted" Data connection between TOM pumps in the
"Infusion taken over by
other pump"
The red LED extinguishes with the acknowledgement of the alarm.
Caution:
pump is in the service mode and is not permitted to be used on a patient. The pump is
then to be checked by a service technician.
If a wrench
is displayed and/or a yellow, red and blue LED blink then the
Pump calibration data have changed (e.g. after an
update).
Recalibrate device via the service program.
The emergency release button was pressed and the
claws manually opened.
Take out syringe and contact technical service department.
plunger plate sensor on the pump.
Check system for negative pressure and eliminate
cause. Consider "Patient Safety".
The set standby time has ended.
Set new time or continue with previously set therapy.
It is not possible to use the pump without a
battery pack.
Turn off pump and insert battery pack according
to describtion "Overview Perfusor® Space".
Therapy and pump settings could not be restored.
Enter therapy and pump settings anew.
Therapy data could not be restored.
Enter therapy anew.
An attempt was made to stop or switch the pump off
without entering the code.
Enter the correct code in order to continue the
therapy or in order to turn the pump off.
SpaceStation has been lost and TOM has been aborted
(Take Over Mode only).
TOM may be reactivated by pressing
when prompted "Return to Take Over M."
Infusion been handed over to second Perfusor® Space
pump (Take Over Mode only)
ALARMS
k
u
and then
59