Pole clamp handle
Pole clamp
release button
Pump slots
4
PERFUSOR® SPACE OVERVIEW
Syringe Fixation
Pull and turn the syringe holder to the right
to open the green axial fixation (see red
arrow). Syringe must be fixed with wings
upright in the slot (found to the left hand
side of the axial fixation) before closing
syringe holder. Make sure that syringe is
properly inserted.
Caution:
Don't touch piston brake when
moving forward.
Fixaton of PoleClamp (Universal Clamp)
Line up bar of pump with bar of PoleClamp
and slide PoleClamp forward until locking
mechanism clicks.
To remove, press release button on frame,
push handle down and pull PoleClamp
backwards.
Transport
A maximum of three pumps (Perfusor®
Space or Infusomat® Space) plus one
SpaceControl may be stacked together (in
ambulance cars or helicopters only one
pump). Avoid external mechanical influence.
Locking Devices Together
Line up the bar of the lower pump with the
bar of the pump above and slide the lower
pump backwards until the lock clicks and
the green buttons are above each other.
To disconnect, push green locking buttons
of top pump device and slide bottom pump
forward.
Caution:
Avoid external mechanical action.
Pole Fixation
Push the opening of PoleClamp against the
vertical pole and lock the screw tightly.
Unscrew to release.
For vertical fixation of PoleClamp push
lever down and rotate either way until lever
clicks into notch. Push lever for rotation.
Caution:
A maximum of three B. Braun
Space pumps can be stacked together
when used with the PoleClamp SP.