Chapter 11
78
WARRANTY / TSC* / SERVICE / TRAINING / DESINFECTING / DISPOSAL
Caution: Do not allow liquids to enter into or come into contact with
any openings or electrical connections on the pump or power supply.
Fluid exposure in these areas may result in the risk of short circuit,
corrosion or breakdown of sensitive electrical components, and/or
electrical shock. If fluid exposure occurs, the device should be swapped
out with another device in a manner that presents minimal interruption
to patient care. The device should remain unplugged until it can be
inspected by a trained technician for any evidence of damage and/or
residual moisture which may impair the function of the device.
Note:
The use of unappoved cleaners and failure to follow the disinfection pro-
cedures and the manufacturer's recommended dilutions can result in an instru-
ment malfunction or product damage and could void the warranty.
Disposal
The pumps as well as battery packs can be returned to B. Braun for further
disposal. When taking care of disposing of disposables as well as infusion soluti-
ons, please consider the applicable hygiene and disposal regulations.
Inspection on Delivery
Despite careful packaging, the risk of damage during transport cannot be
entirely prevented. Upon delivery, please check that all items are present. Do not
use a damaged device. Contact the service department.
Testing the proper funciton of the device should be performed before initial use.
This is even ruled by law in several countries. A respective form can be obtained
from B. Braun.
Included in Delivery
Perfusor® Space, Battery-Pack SP, Instructions for Use-Set.