Chapter 3
Example:
Program Mode should only be performed by an experienced user being familiar with
the principles of the Program Mode function and properly trained in using the present
device.
Note:
(
Note:
Set Profile Parameters: The therapy can be started directly via the drug library or via
the Main Menu/Special functions.
Starting Program Mode via Drug Library:
Note:
before and have been uploaded into the pump.
• Switch on pump with
• Insert disposable and use the drug library according to Instructions for Use.
• Select the desired drug with
• Select Program Mode with l.
In the following screen the user has to confirm the number of steps for the
therapy with k.
The settings for the steps of the infusion are shown on the display. These
settings, configured in the Drug List Editor, need to be confirmed with r.
• To change the values, press
• Adjust VTBI with q.
The active Program Mode function always displays this icon in the Display
).
Bolus function is disabled for Program Mode.
Program Mode settings have been configured in the Drug List Manager
o
and wait until self-check is finished.
t
and press l.
l
to change and
SPECIAL FUNCTIONS
k
to confirm.
35