SPECIAL FUNCTIONS
Chapter 3
Recommendation
Ensure first Perfusor® Space pump is infusing
Ensure first Perfusor® Space pump must be
running in 'continuous mode' (i.e. ml/h or a
dose rate; not KVO, PCA etc.)
Deactivate Data Lock
Data connection must be active between
pumps – check the positioning of pumps in
the SpaceStation
Ensure both pumps must have a syringe of
the same size
Ensure both pumps have the same Care Unit
selected
Ensure same patient profile selected on both
pumps
Both pumps must have the same software
version – contact your service department
Both pumps must have the same modification
data – contact your service department
Both pumps must have the same 'Stop at syr.
end' settings – contact your service
department
Both pumps must have TOM activated –
contact your service department
More information regarding alarm hints may be found in chapter 5.