Chapter 3
In this state the patient is allowed to demand boli. Depending on the status of
the therapy these are either administered or denied. Changing the syringe is also
possible by using the code for level 1 or level 2. Altering the settings for PCA or
other therapies however is only possible with the code for level 3.
The status of the therapy can be checked in the menu „Status" .
Enter the „Main Menu" with
An acoustic confirmation of demanded boli can be activated and modulated by
t
in Data Lock 3.
Is a demand button connected, the therapy symbol looks like this:
In case there is no demand button connected the therapy symbol looks like
this:
The demand button is connected to the interface P2 at the rear side of the
pump.
Hint:
PCA later on (in case the drug is dedicated for use with continuous and PCA
application).
SpacePCA-Chart
r
If
The bar represents a time axis, with the points above the axis representing the
number of boli administered and the points below the axis representing the
number of boli refused.
The chart has a 15 minute resolution and shows max of 5 points per 15 min-
utes. Should more then 5 boli be given or refused in this time, the last point will
be turn bold.
Changes to the PCA parameters are displayed as arrowheads at the bottom of
the chart.
.
It is possible to start a therapy in continuous mode and switch over to
is pressed on the RUN screen, the SpacePCA-Chart is displayed:
c
and select the "Status" with q.
The A/D-ratio indicates the
percentage of administered and
demanded boli thus giving an idea
about the effectivity of the therapy.
SPECIAL FUNCTIONS
25