9 Technical Data and Conformity  |  9.1 Power Requirements and Operating Data
Tube 30°/20 with int. 3 MP
Primotech  |  Document 430055-9044-008 V.1.0; 03 / 2014
Property
camera
Viewing angle
Viewing height
Interpupillary distance
Optical split ratio
Camera adapter
magnification
Camera field of view,
diagonal
Sensor
Live / Video recording
over LAN / WLAN
Snap resolution
Auto white balance
Electrical interfaces
Buttons
Value
30°
380 - 415 mm
Adjustable, 48 - 75 mm
50% / 50%
0.39x
73% of eyepiece field of view
(sensor cropped for performance reasons)
Micron MT9P031, 1/2.5" (7.13 mm diag.)
CMOS, 24 bit color, 2560 x 1920 pixels, 2.2 μm
pixel size
Spectral sensitivity without IR filter 400 - 700 nm
¢
640 x 480 pixels (VGA)
¢
20 fps, latency: ~400 ms
¢
Bitrate: 1.5 / 3 / 6 Mbit/s
3 MP / 2048 x 1536 pixels, YUV color
Yes (Auto/Lock)
¢
12 V DC power input (provided by the
microscope via interconnection cable)
¢
Network (RJ45), 100 Mbit/s
Reset button (backside)
53