Technical Data And Conformity; Power Requirements And Operating Data - Primotech MAT Instruction Manual

Upright microscope
Table of Contents

Advertisement

9 Technical Data and Conformity  |  9.1 Power Requirements and Operating Data
Primotech  |  Document 430055-9044-008 V.1.0; 03 / 2014

9 Technical Data and Conformity

9.1 Power Requirements and Operating Data

The Primotech system must be connected to the AC network via the central power
supply by a country-specific mains power cable.
Property
Protection class
Degree of protection
Electrical safety
Overvoltage category
Radio interference
suppression
Resistance to interference
Line voltage for power
supply
Line frequency
Power consumption of
power supply (plug)
Power supply (plug)
output
Microscope 12 V /
6 V DC
LED class of complete
device
Value
II
IP 20
In accordance with DIN EN 61010-1 (IEC 61010-1)
including CSA and UL regulations
II
In accordance with EN 55011
In accordance with DIN EN 61326-1
100 to 240 V ±10%
The supply voltage does not need to be
transformed.
50 to 60 Hz
Max. 70 VA; secondary voltage of external power
supply 12 V
12 V DC; max. 2.5 A
Adjustable from 1.5 V to 6 V
LED risk group 1 according to DIN EN 62471