6 Changing Components  |  6.4 Replacing the Light Source
Procedure 1
Primotech  |  Document 430055-9044-008 V.1.0; 03 / 2014

6.4 Replacing the Light Source

If the transmitted illumination light source stops working, the entire light source
needs to be replaced, even if only the bulb has stopped working.
INFO
The reflected illumination light source cannot be changed.
To change the transmitted illumination light source:
Use the 1.5 mm hex key to loosen the two hexagonal bolts of the light source
in the direction of the arrows.
2
Grip the handle of the light source and slide it out of the stand.
3
Slide in the replacement light source.
Use cotton gloves to avoid fingerprints on the new light source.
4
Use the 2.5 mm hex key to tighten the two hexagonal bolts.
43