Model 2700 Multimeter/Switch System User's Manual
Table H-2 (continued)
LabVIEW examples
Name
Simple3
None
Simple4
None
Manual
Reference
Use Case 3 — Two scans using 7708 module:
• 40 channel DCV (1V range) scan.
• 20 channel Ω4 scan:
• Models 2700 and 2701 – 100Ω range.
• Model 2750 – 10Ω range, dry-circuit ohms enabled.
• Measurement speed (rate) – 0.1 plc.
• DCV input divider – Enabled (10MΩ input impedance).
• Filter – Disabled (no filtering).
• Buffer – Store 40 DCV reading strings, 20 Ω4 reading strings.
Buffer elements include reading only.
• Limits (DCV scan) – Limit 1 (all channels) = 20mV, Master
Latch enabled.
• Triggering – Bus control source.
Use Case 4 — Two scans using 7708 module:
• 40 channel DCV scan (1V range). Configuration saved in User
Setup 1.
• 20 channel Ω4 scan. Configuration saved in User Setup 2.
• Models 2700 and 2701 – 100Ω range.
• Model 2750 – 10Ω range, dry-circuit ohms enabled.
• Setup 1 or Setup 2 recalled to perform scan.
• Measurement speed (rate) – 0.1 plc.
• DCV input divider – Enabled (10MΩ input impedance).
• Filter – Disabled (no filtering).
• Buffer – Store 40 DCV reading strings, 20 Ω4 reading strings.
Buffer elements include reading only.
• Limits (DCV scan) – Limit 1 (all channels) = 20mV, Master
Latch enabled.
• Triggering – Bus control source.
KE2700 Instrument Driver Examples
Brief Description
H-17