red. This is normal as the system has not yet been fully commissioned and this can be ignored at
this stage.
3. The installation wizard will start automatically. The wizard will take you through the process of
connecting all the gas inputs, filling the antifoam cartridge (refer also to
preparation
on page 20), EC sensor cartridge replacement, antifoam pump maintenance and
the instrument auto-check (refer also to
Post-installation
1. The instrument has been delivered with English as the default language. However, a USB
memory stick is supplied with other available languages (German, Spanish, Chinese, and
Japanese). If you wish to keep English as the preferred language please proceed to the next step
(Step 2 below), otherwise follow these instructions for installing a new language:
• Set the instrument power to OFF.
• Insert the supplied USB memory stick into the USB port on the instrument.
• Set the instrument power to ON.
• Follow the on-screen instructions to install the preferred language.
• Remove the USB memory stick and restart the instrument.
2. Set the security levels, user ID's and passwords as described in
on page 13.
3. Configure the instrument parameters as described in
4. Perform a barometric sensor calibration as described in
on page 16.
5. Perform an O
sensor calibration as described in
2
residual value of the sensor is low enough (controlled automatically by the instrument). This
operation may take around 1-2 hours depending on storage conditions.
Note: The other sensors are more stable and will not require calibration prior to use.

Power supply

A power cable can be ordered with the instrument and this cable must be used to connect the instrument to the
power supply. If no power cable was ordered, please contact your Hach Lange representative for cable
specifications. Inadequately dimensioned power cables must not be used.
Check voltage requirement sticker on the rear panel of the instrument before connecting (100-240 VAC).
Connect the power cable to the socket at the rear of the instrument (left in diagram) for AC
power connection. To switch the instrument on and off, press the rocker switch (right in the
diagram) - I for On and O for Off.
Note: For optimum performance always keep the gas and power supplies to the instrument switched on.
Emergency STOP button
The red STOP button is located on the right side of the instrument. Should the need arise to stop the
machine at any time during operation, press this button. An audible click can be heard as the button
locks in position. The instrument should then be disconnected from the power supply. Once
disconnected, unlock the button by turning it clockwise. The instrument can then be reconnected to
the power supply and restarted.
Instrument auto check
Configuration
Barometric pressure sensor
Oxygen sensor
W A R N I N G
C A U T I O N
Antifoam cartridge
on page 8).
Security and user management
on page 9.
on page 17. Wait until the
English 7