Message
E10: Foam has been detected in the
gas path. The measurement process
is canceled.
E20: Not enough pressure in the
system. The measurement process is
canceled.
E40: Check gas supply, current
pressure: nn
E80: No message displayed.
E100: Ultrasound error.
E200: Motor error - High limit.
E800: Measurement stopped. The
needle pressure is too low.
E1000: Measurement stopped.
Headspace pressure is not stable.
E2000: No message displayed.
E4000: Calibration fails. A leak during
the headspace determination
prevents a correct measurement.
E10000: The measurement could not
start. The rim detector is faulty.
24 English
Probable cause
There is a problem with the
antifoam injection.
Incorrect tilt of the sonotrode.
There is no more antifoam.
The pressure in the package is
greater than the purge gas
pressure.
Problem of flow regulation during
the measurement.
The gas path is partially or totally
blocked.
Before filling the reference volume,
the pressure is lower than 1.5 bar.
No package installed.
The ultrasound power supply is
OFF.
The emergency STOP button is not
released.
There is too much water in the
ultrasound generator.
The motor is above the reference
position.
The needle pressure is too low,
probably caused by air leaking
from the piercing position.
The headspace pressure is
unstable.
Measurement aborted by user.
Bad piercing position.
The detector has become clogged
with old sample and dirt.
Solution
Try another measurement.
Check the tilt of the sonotrode.
Check the antifoam level and refill if
necessary.
Prime the antifoam pump.
Increase the purge gas pressure.
Check the purge gas pressure.
Change the Goretex filter.
Control the antifoam injection.
Check the purge gas pressure.
Install a package.
Check the ultrasound power supply.
Release emergency STOP button
by turning it clockwise.
Remove any excess water.
Go to the maintenance menu and
initialize it.
Pierce the package in a different
position.
Change the Goretex filter if
clogged.
Check correct piercing tip is being
used for the package (PET/metal).
Check for package leaks.
N/A
Pierce the package in a different
position.
Change the Goretex filter.
Clean the two metal rods either
side of the piercing device with a
damp cloth. Turn the instrument off
and restart it.
Check the connections.
Change the rim detector.