OCULUS Pentacam Interpretation Manual page 85

High-resolution rotating scheimpflug camera system for anterior segment analysis
Hide thumbs Also See for Pentacam:
Table of Contents

Advertisement

The Pentacam® uses a technique of optical cross-sectioning to identify the anterior and posterior
corneal surface, the anterior iris, and the anterior and posterior surface of the lens
measuring these surfaces and their relative position elevation maps of the anterior and posterior
cornea can be produced as well as a full corneal thickness map (additionally lens measurements,
but these will not be discussed here).
Figure 100: Scheimpflug image
Corneal thickness maps are easy to comprehend. Corneal thickness is an absolute measurement
representing the distance between the anterior and posterior surface. Corneal thickness is reference
independent. Elevation is different. Elevation is measured by locating points in space and is
displayed relative to a reference surface. Raw elevation data is displayed (measured) against a
planar reference surface. This represents the "raw" elevation data
Figure 101: Raw elevation data
12 Belin/Ambrósio Enhanced Ectasia Display
(Figure
(Figure
101).
100). By
83