Screening For Refractive Surgery By Prof. Michael W. Belin; Screening Parameters, 4 Maps Refractive Display; Suggested Installation Settings - OCULUS Pentacam Interpretation Manual

High-resolution rotating scheimpflug camera system for anterior segment analysis
Hide thumbs Also See for Pentacam:
Table of Contents

Advertisement

10
Screening for refractive surgery
by Prof. Michael W. Belin
10.1

Screening parameters, 4 Maps Refractive Display

10.1.1 Suggested installation settings

The following are my guidelines for pre-operative refractive surgery screening for keratoconus:
Use the 4 Maps Refractive Display showing anterior elevation, posterior elevation, pachymetry
and anterior sagittal curvature. It is advisable to keep the display, scales and colors constant for
refractive screening, as this will allow for a rapid visual inspection
Pachymetry
Right-click on the scale and set Abs: normal, (300-900 μm)
Right-click on the actual display to open the drop down menu. Turn on the following: Show Apex
Position (1), Show Thinnest Location (2), Show Pupil Edge (6), Show Nasal/Temp (7), Show Max
Diameter 9mm (12) and Show Numeric Values (14)
Anterior elevation & posterior elevation
Right-click on the scale and set to
"Belin intuitive" +/- 75 μm for refractive practice
"Belin intuitive" +/- 150 μm for medical practice
Best-fit-sphere (BFS), float, manual, BFS diameter set to 9.0 mm or 8.0 mm
On the 9.0 mm display you should have no or minimal extrapolated data for the study to be valid
Right-click on the display and turn on the following: Show Apex Position (1), Show Thinnest
Location (2), Show Pupil Edge (8), Show Nasal/Temp (9), Schow Max Diameter 9mm (14) and Show
Numeric Values (15)
Sagittal curvature
Right-click on the scale and set to Abs: Normal, American Style and Diopter
Right-click on the display and set to Show Min. Radius Pos. Front (3), Show Pupil Edge (6), Show
Nasal/Temp (7), Schow Max Diameter 9 mm (12) and Show Numeric Values (13)
Note:
The different borderline numbers for the elevation maps depend on the BFS diameter you are using,
i.e. 9 mm or 8 mm.
10 Screening for refractive surgery
47

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Pentacam hr

Table of Contents