FEATURES AND FUNCTIONS
Functions setting
(1) Inflation level setting
P - SET
AUTO (Automatic setting): Can be used when the SINGLE, AVG, or MANU Mode is selected.
The unit estimates the systolic blood pressure during inflation and inflates to a proper cuff pressure
(approximately 30-40 mmHg above the patient's systolic pressure).
Manual level setting: Inflation level can be set manually between 100 and 280 mmHg. Set the
level to 30 to 40 mmHg higher than the expected systolic pressure.
• To set the P-SET to "AUTO," turn the P-SET Knob counterclockwise as far as it goes until you can hear a click.
• In the "AUTO" setting, inflation level may not be set automatically when the systolic blood pressure is more than
220 mmHg. Use the unit in the manual setting.
• If the cuff has not been inflated to the necessary level, it may be re-inflated automatically.
HIDE
(2) Non-display function
Use to prohibit the display of measurement results. However,
the cuff pressure during measurement is displayed.
This function can be used in the SINGLE and AVG Modes.
HIDE
By pushing the
Button, display or non-display of status
is switched alternatingly.
SYS
mmHg
Display
Non-display
DIA
mmHg
PULSE
BPM
AC
(3) Manual deflation control
Accelerate deflation by pushing the DEFLATION (deflation
control) / Measurement Result Display Switch Button during
deflation in the measurement by the MANU Mode.
With each push of the
Button, cuff is deflated rapidly
in increments of 5 to 10 mmHg.
ON/OFF
START
HIDE
P-SET
MODE
SINGLE
140
180
MANU.
AVG.
220
CHECK
100
STOP
260
AUTO
280
AC
ON/OFF
START
HIDE
P-SET
MODE
SINGLE
140
180
AVG.
MANU.
100
220
CHECK
STOP
260
AUTO
280
12
FEATURES AND FUNCTIONS
List of Functions for each Mode
Measurement
Single (SINGLE)
Mode
Mode
Function
Inflation level set-
ting
Non-display func-
tion
Manual deflation
control
(4) AVG Function setting
You can set the number of measurements, the waiting time until the 1st measurement, and the
measurement interval for the AVG Mode.
Items to set
Function #
F1
Number of measurements
F2
Waiting time until the start of 1st measurement
F3
Measurement interval
Note: The bold letters represent the factory-set values.
Procedure to change the set values
1) When the power is OFF, push the ON/OFF (power) Button
for more than three seconds while holding the START Button;
F1 is displayed.
2) Push the START Button and select the function to set from
F1 to F3. Each time you push the START Button, the func-
tions change in the order of
3) Push the DEFLATION (deflation control)/Measurement
Result Display Switch Button and change the set values.
4) When the setting is finished, push the ON/OFF (power)
Button to turn off the power. The setting is changed.
Average
Auscultation
Check
(AVG) Mode
(MANU) Mode
(CHECK) Mode
Set value
2 times or 3 times
0 sec, 3 min, 5 min, or 10 min.
5 sec, 30 sec, 1 min, or 2 min.
XL
ON/OFF
AC
START
HIDE
P-SET
MODE
140
180
SINGLE
AVG.
MANU.
220
CHECK
100
STOP
260
AUTO
280
XL
.
ON/OFF
AC
START
HIDE
P-SET
MODE
SINGLE
140
180
AVG.
MANU.
220
CHECK
100
STOP
260
AUTO
280
XL
min
ON/OFF
sec
AC
START
HIDE
P-SET
MODE
140
180
SINGLE
AVG.
MANU.
220
CHECK
100
STOP
260
AUTO
280
13

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents