After several seconds the PRNU calibration will end and the correction coefficients will be
enabled. The system gain remains as first set.
The coefficients and gain parameters, timing and control configuration can be stored in any
one of eight user sets and automatically retrieved at power-up or by user selection.
CamExpert has a default timeout of 20 seconds per command, which is too
short for the FFC calibration to run fully. You can change the default timeout by
setting a command line argument in the short-cut:
Right-click on the short-cut in the start menu and select Properties.
In the Target field, add the switch –timeout 60 (See below)
This increases the command timeout to 60 seconds.
Note that you must include a character space between the closing
quotation mark in the target and the hyphen before the timeout value.
Repeat for desktop short-cut

Flat Field Parameters

This Flat Field category contains a number of features that are used to correct image distortion due
to lens vignetting and uneven illumination.
30 • Camera Operation