Omega Engineering OMEGAFLEX FPU5-MT-110 User Manual page 62

Peristaltic pump motor
Table of Contents

Advertisement

Error Message
Locked Rotor?
(continued)
Lost PHASE SYNC!
Must SET RPM!
Must SET TIME!
.
Problem
B
There may be an internal problem.
There may be an internal problem.
A
You have attempted to pump in
the RPM Mode with the RPM
set to 0 rpm.
B
You have attempted to dispense
fluid during calibration with the
RPM set to 0 rpm.
You have attempted to
pump in the TIME
Mode with the TIME set
to 00:00:00.
Troubleshooting Guide
Solution
B
For internal problems, call Customer
Service.
1.
Turn the unit's power off and
on, adjust the settings, and
attempt to resume pumping.
2.
If the problem persists, call
Customer Service.
A
1.
Press the RPM key.
2.
Use the UP and DOWN
ARROW keys to adjust the
RPM setting.
3.
Press the START/STOP key to
start the pump.
B
1.
Press the RPM key.
2.
Use the UP and DOWN
ARROW keys to adjust the
RPM setting.
3.
Press the CALIBRATE key to
restart the calibration.
1.
Press the TIME key.
2.
Toggle from SECONDS to
MINUTES to HOURS with the
TIME key.
3.
Use the UP and DOWN
ARROW keys to adjust the
TIME setting.
4.
Press the START/STOP key to
start the pump.
7
7-3