Selecting Tubing; Tubing Life; Loading The Tubing - Omega Engineering OMEGAFLEX FPU5-MT-110 User Manual

Peristaltic pump motor
Table of Contents

Advertisement

3
Getting Started

3.2 Selecting Tubing

Select a tubing material and size that is right for your application (the fluid and
flow rate that you are pumping).
Table 9-2 in Chapter 9, shows the average flow rates for different size tubing.
Normalized flow rates (mL per revolution) vary significantly, based on motor
speed, tubing materials, viscosity, and mechanical tolerances in pump dimensions.
Table 9-3 outlines the variances resulting from differences in motor speed, tubing
materials, and mechanical tolerances. Table 9-6 outlines the variances due to
differences in viscosity.
To determine the chemical compatibility of a particular tubing material, it is
recommended that you test the tubing under actual conditions.
Tubing materials that can be used include Vinyl, Viton, Tygon, Silicone,
Santoprene, and Norprene. Up to 68 durometer tubing can be used.

3.3 Tubing Life

Over time and high speeds, flow rates will drop as the tubing wears out. Tubing
life for various materials and sizes are shown in Table 9-4 in Chapter 9. Tubing
should be periodically inspected for wear. Either move the tubing to a fresh
section, or replace tubing entirely (refer to Section 3.4).

3.4 Loading the Tubing

Follow this procedure (refer to Figures 3-8 through 3-12):
Stator
Latch
Figure 3-8. Parts of the Pump
3-4
Poor tubing life results were obtained for
Santoprene tubing. This particular tubing should not be used with the
pump motor.
Perform ALL of the steps in Chapter 3 before turning the motor on.
Clamp
Screw
Clamp
Plate
Tubing
Figure 3-9. Opening the Stator
NOTE
" (8.0mm) inner diameter
5
1 6
CAUTION
Figure 3-10. Loading the Tubing

Advertisement

Table of Contents
loading

This manual is also suitable for:

Fpu5-mt-220