Set Up Scales; Operation; Weighing; Start Weighing - Seca 374 User Manual

Hide thumbs Also See for 374:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 31

4.2 Set up scales

5. OPERATION

5.1 Weighing

Start weighing

40 •
The scales are fully assembled on delivery.
ATTENTION!
Incorrect measurement with force shunt
If the housing of the scales is resting on some-
thing, e.g. a towel, the weight measurement will
not be correct.
− When setting up the scales, ensure that they
only make contact with the ground via the
foot screws.
1. Place the scales on a flat, stable surface.
2. Align by turning the foot screws.
WARNING!
Serious injury to baby following a fall
Baby scales generally stand on raised work sur-
faces. If the baby falls from this surface, this
may result in serious irreversible or fatal injuries.
− Never leave a baby unsupervised on the
scales.
1. Check that there is no load on the scales.
2. Press the Start key.
The display will read
through all elements of the display.
The scales are ready for operation when the display
0. 0 00
reads
.
If the device is being operated with a power supply
unit, the symbol
3. Place the baby on the scales.
4. Read off the measured result.
se[A
and then quickly run
will appear in the display.

Hide quick links: