Adjustment Of Scales [374|376|378|703|704|634|635|644|645|656|657|664|665|674|675| 676|677|684|685|264|274|284|285|954|957|959|963]; General[374|376|378|703|704|634|635|644|645|656|657|664|665| 674|675|676|677|684|685|264|274|284|285|954|957|959|963] - Seca 6641321108 Service Manual

Table of Contents

Advertisement

Maintenance
Firmware check digit
Every firmware is given its own check digit.
Device check digit
The overall check digit for the device is the total of the individual check digits.
Representation on display
Hardware module
Firmware version with index / internal seca value
PTB version number
PTB check digit
Weight module
290.08
01.08
1A47
M704
290.08 / 290.08
02.16
348E
2.1
Adjustment of Scales
[374|376|378|703|704|634|635|644|645|656|657|664|665|674|675|
676|677|684|685|264|274|284|285|954|957|959|963]
2.1.1 General[374|376|378|703|704|634|635|644|645|656|657|664|
665|674|675|676|677|684|685|264|274|284|285|954|957|959|963]
To compensate for linear measuring deviations, which occur e.g. as a result of
gravity variations in different gravity zones, the scale offers an adjustment
feature. This adjustment must also be carried out whenever the load cell is
replaced.
14.11.12/RLEU/ARI/MRE
Display
290.08
01.08
1A47
10/40
Compact
Weight module
290.08
01.08
1A47
M703
290.08
01.08
1A47
seca
Extended display
290.08
298.08
01.08
01.08
1A47
M285
290.08 / 298.08
02.16
5CAD
30-34-00-812c
4266

Advertisement

Table of Contents
loading

Table of Contents