The SpO
and Pulse
2
Rate
cannot
be
displayed normally
The SpO
and Pulse
2
Rate
are
not
displayed stably
The device cannot
be turned on
The display is off
suddenly
The device cannot
be used for full time
after charge
The battery cannot
be full charged even
after
10
hours
charging time.
9 Key of Symbols
Signal
PRbpm
CMS-60C
Medical Test Supply
1. The finger is not properly positioned.
2. The patient's SpO
2
detected.
1. The finger is not placed inside deep
enough.
2. The finger is shaking or the patient is
moving.
1. The battery is drained away or almost
drained away.
2. The battery is installed incorrectly.
3. The malfunction of the device.
The battery is drained away or almost
drained away .
1.The battery is not full charged.
2. The battery is broken
The battery is broken
Warning – See User Manual
The pulse oxygen saturation(%)
Pulse rate (bpm)
Close the alarm sound indication forever
Pause the alarm sound indication
Open the alarm sound indication
Close the pulse sound indication
1. Place the finger properly and
try again.
is too low to be
2. Try again; Go to a hospital for
a diagnosis if you are sure the
device works all right.
1. Place the finger properly and
try again.
2. Let the patient keep calm
1. Please recharge the battery
2. Please Install the battery again.
3. Please contact the local service
center.
Please recharge the battery
1. Please recharge the battery
2. Please contact the local service
center.
Please contact the local service
center.