G. Strenuous action of the subject or extreme electrosurgical interference may also affect the
accuracy.
H. Testee cannot use enamel or other makeup.
I. Please clean and disinfect the device after operating according to the User Manual(7.1).

6.3 Clinical restrictions

A. As the measure is taken on the basis of arteriole pulse, substantial pulsating blood flow of subject
is required. For a subject with weak pulse due to shock, low ambient/body temperature, major
bleeding, or use of vascular contracting drug, the SpO
waveform (PLETH) will decrease. In this
2
case, the measurement will be more sensitive to interference.
B. For those with a substantial amount of staining dilution drug (such as methylene blue, indigo
green and acid indigo blue), or carbon monoxide hemoglobin (COHb), or methionine (Me+Hb) or
thiosalicylic hemoglobin, and some with icterus problem, the SpO
determination by this monitor
2
may be inaccurate.
C. The drugs like dopamine, procaine, prilocaine, lidocaine and butacaine may also be a major factor
blamed for serious error of SpO
measure.
2
D. As the SpO
value serves as a reference value for judgment of anemic anoxia and toxic anoxia,
2
some patients with serious anemia may also report good SpO
measurement.
2

7 Maintain, transportation and storage

7.1 Cleaning and Disinfecting

Using medical alcohol to disinfect the device, nature dry or clean it with clean soft cloth.

7.2 Maintain

A. Please clean and disinfect the device before using according to the User Manual(7.1).
B. Please recharge the battery when the screen shows
.
C. Recharge the battery soon after the over-discharge. The device should be recharged every six mo
nths when it is no regular used. It can extend the battery life following this guidance.
D. Please take out the batteries if the oximeter is not in use for a long time.
E. The device needs to be calibrated once a year (or according to the calibrating program of
hospital). It also can be performed at the state-appointed agent or just contact us for calibration.

7.3 Transportation and storage

A. The packed device can be transported by ordinary conveyance or according to transport contract.
The device cannot be transported mixed with toxic, harmful, corrosive material.
B. The packed device should be stored in room with no corrosive gases and good ventilation.
Temperature: -40°C~60°C; Humidity: ≤95%

8 Troubleshooting

Trouble
Possible Reason
Solution