Sensitivity adjustment 1602 MP 8 using an
external calibration weight. Remove cap from
calibration button (4). Tare (6) and wait for stability
before pressing calibration button (4) Display
shows "C" and the busy symbol (21). Load
calibration weight (150 g, tolerance ±0.1 mg).
Wait for display to show "150.0000 g" and for
beeper. Balance returns to weighing mode
automatically.
Sensitivity adjustment 1612 MP 8 with the
built-in calibration weights
1. Semimicro range (30 g/0,1 mg)
Clear pan. Tare (6).
Wait for stability (7).
Slightly depress operating mode switch (3) and
carefully set it to "Cal 30 g" (see picture 7d).
Display shows "C" and the busy symbol (21). lf
you get CE display, restart calibration from
tare(6). Place 30 g calibration weight (tol. ±0.02
mg) on pan. When CC is displayed, balance will
automatically return to weighing function.
2. Macro range (160 g/0.1 mg)
Tare (6) and wait for stability (7).
Slightly depress operating mode switch and set it
carefully to "Cal 160 g" (see picture 7c). Display
shows "C" and the busy symbol (21).
(lf CE is displayed, restart calibration from stand-
by position
).
When CC appears in display, reset switch (3) to
position "l".
9