Sartorius 1601 MP 8 Installation And Operating Instructions Manual page 8

Electronic analytical balances

Advertisement

Busy-Symbol (21) signals temporary power
failure (disappears when you push tare button (6)
or indicates that microprocessor is completing
previous function and not ready for the new job..
Place 100 g calibration weight on pan.
Waitfordisplay"100.0000g"and the beeper signal.
Balance automatically returns into weighing
program.
8
Sensitivity adjustment 1601 MP 8
is done with external calibration weight only
(see accessories 100g ± 0.15mg).
Remove cap from calibration button (4) clear pan,
tare (6), and wait for stability (7). Press calibration
button (4). Display shows "C" and the busy
symbol (21) (if display shows "CE", repeat
calibration starting from tare.)
Sensitivity adjustment 1602MP8 using the
built-in calibration weight.
Tare (6). At stability (7), set operating mode
switch (3) to "CAL". Display shows "C" and the
"Busy Symbol" (21). (lf "CE" appears, restart
adjustment with the "Stand-by" position "0"-) Wait
for display to show CC and beeper Signal before
you set operating mode switch (3) back to
position "l".