Trigger; Table 3-2. Actual Range And Measurement Precision - National Instruments DAQ PCI-4451 User Manual

Dynamic signal acquisition device for pci
Hide thumbs Also See for DAQ PCI-4451:
Table of Contents

Advertisement

Chapter 3
Hardware Overview
Note
The device boots in a mode with the outputs disabled
Although these functions appear similar, they are quite distinct and are
implemented to protect your external equipment from startup transients.
When the DACs no longer have data written to them, they automatically
retransmit the last data point they received. If you are expecting the data to return
to 0 V or any other voltage level, you

Trigger

PCI-4451/4452 User Manual

Table 3-2. Actual Range and Measurement Precision

Attenuation
Linear
1.0
0 dB
10
20 dB
100
40 dB
1
The value of 1 LSB of the 16-bit DAC; that is, the voltage increment corresponding to a
change of one count in the DAC 16-bit count.
See Appendix A, Specifications, for absolute maximum ratings.
All data written to the DACs are interpreted as two's complement format.
In two's complement mode, data values written to the analog output
channel are either positive or negative.
In addition to supporting internal software triggering and external digital
triggering to initiate a data acquisition sequence, the PCI-4451/4452 also
supports analog level triggering. You can configure the trigger circuit to
monitor any one of the analog input channels to generate the level trigger.
Choosing an input channel as the level trigger channel does not influence
the input channel capabilities. The level trigger circuit compares the full
16 bits of the programmed trigger level with the digitized 16-bit sample.
The trigger-level range is identical to the analog input voltage range. The
trigger-level resolution is the same as the precision for a given input range.
Refer to Table 3-1.
The trigger circuit generates an internal digital trigger based on the input
signal and the user-defined trigger levels. Any of the timing sections of the
DAQ-STC can use this level trigger, including the analog input, analog
output, RTSI, and general-purpose counter/timer sections. For example,
Attenuation
dB
±10.0 V
±1.00 V
±0.100 V
dB
0 V
append the data to make it do so.
MUST
3-6
Range
Precision
305.18 µV
30.158 µV
3.0518 µV
0 V
infinitely ( ) attenuated.
AND
© National Instruments Corporation