Differential Connection Considerations (Diff Input Mode) - National Instruments PXI-7811 User Manual

Multifunction rio
Table of Contents

Advertisement

Chapter 2
Hardware Overview of the NI 78xxR

Differential Connection Considerations (DIFF Input Mode)

In DIFF input mode, the NI 783xR/784xR/785xR measures the difference
between the positive and negative inputs. DIFF input mode is ideal for
measuring ground-referenced signals from other devices. When using
DIFF input mode, the input signal connects to the positive input of the
instrumentation amplifier and its reference signal, or return, connects to the
negative input of the instrumentation amplifier.
Use differential input connections for any channel that meets any of the
following conditions:
Differential signal connections reduce noise pickup and increase
common-mode noise rejection. Differential signal connections also
allow input signals to float within the common-mode limits of the
instrumentation amplifier.
R Series Multifunction RIO User Manual
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
The input signal is low level (less than 1 V).
The leads connecting the signal to the NI 783xR/784xR/785xR are
greater than 3 m (10 ft).
The input signal requires a separate ground-reference point or return
signal.
The signal leads travel through noisy environments.
2-10
ni.com

Advertisement

Table of Contents
loading

Table of Contents