Specifications - National Instruments Isolated Analog Output Module SCC-AO10 User Manual

National instruments isolated analog output module user guide
Hide thumbs Also See for Isolated Analog Output Module SCC-AO10:
Table of Contents

Advertisement

Specifications

Electrical
© National Instruments Corporation
5.
Adjust the offset potentiometer on the top of the SCC-AO10 until the
voltmeter reads 0 VDC ±100 μV. The offset is now calibrated.
6.
Using your software, have AO (X) output 9.9 VDC.
Adjust the gain potentiometer on the top of the SCC-AO10 until the
voltmeter reads 9.9 VDC. The gain is now calibrated.
These ratings are typical at 25 °C unless otherwise stated.
Number of output channels .................... One nonreferenced (floating)
Input range ............................................. ±10 V
Output range........................................... ±10 V
Current drive .......................................... ±30 mA
Gain nonlinearity.................................... 0.5% of full-scale output range
Output noise ........................................... 2.5 mV
Bandwidth .............................................. >23 kHz
Slew rate................................................. 1 V/μs
Short circuit protection
Current ............................................ 37 mA max
Duration .......................................... Unlimited
Load impedance ..................................... 10,000 pF max
7
SCC-AO10 Isolated Analog Output Module User Guide
typ; 4 mV
max
rms
rms