ALARM/WARNING:
SYMPTOM
BREATH RATE OVER CAPACITY
ALARM:
Intermittent audible beep and illuminated red
alarm indicator.
Audible beeps and red indicator blinks become
more rapid as breathing rate increases.
SYSTEM TOO HOT/COLD TO START ALARM:
Continuous audible beep and red alarm
indicator is illuminated. Fan is On.
AND
Flow indicators 1 & 2 are illuminated.
SYSTEM RUNNING TOO HOT/COLD ALARM:
Continuous audible beep and red alarm
indicator is illuminated. Fan is On.
AND
Flow indicators 1 & 3 are illuminated.
1148112-F~12
PROBABLE CAUSE
Your breathing rate has exceeded the capacity
of the XPO
concentrator.
2
Concentrator is too hot, or too cold, to allow
it to turn On.
Concentrator has become too hot, or too
cold, during operation.
SOLUTION
1.
Immediately reduce your activity level to
get your breath rate to slow down, and/or
2.
If the alarm continues, change to another
source of oxygen.
1.
Allow concentrator to cool down to less
than 95ºF (35°C), or warm up to 50ºF
(10°C), before turning the concentrator
Off, then On again.
2.
If the alarm continues, change to another
source of oxygen and contact your
equipment provider.
1.
Allow concentrator to cool down to less
than 95ºF (35°C), or warm up to 50ºF
(10°C), before turning the concentrator
On again.
2.
If the alarm continues, change to another
source of oxygen and contact your
equipment provider.