Invacare® XPO₂™
ALARM/WARNING:
SYMPTOM
CHARGING INDICATOR:
With the concentrator turned Off and with
the AC or DC charging adapter plugged into
the concentrator, no Battery Gauge segments
are illuminated when battery status button is
pressed. Battery Gauge is dark.
LOW BATTERY WARNING:
There is a single audible beep and the green
25% battery indicator is blinking.
LOW BATTERY ALARM:
There is a single audible beep and the green
25% battery indicator is rapidly blinking.
BATTERY DISCHARGED ALARM:
There is a triple audible beep and the green
25% battery indicator is blinking very rapidly.
34
PROBABLE CAUSE
Power source is no good, or there is a loose
connection.
Internal Battery is outside the allowed
temperature range for charging.
For DC adapter only, blown fuse.
Internal battery requires charging.
Internal battery requires charging.
Internal battery is completely discharged and
needs to be recharged.
SOLUTION
Try another power outlet and check
connections to charger and concentrator.
Allow unit to reach temperature range for
charging (between 41ºF and 95ºF).
Change fuse. Use Invacare part number
1185420 as fuse for DC adapter.
Connect XPO
concentrator to its AC or DC
2
power adapter, or to its supplemental battery
pack.
1.
Connect XPO
concentrator to its AC or
2
DC power adapter, or to its supplemental
battery pack.
2.
If no other power source is available,
change to a reserve oxygen supply.
1.
Connect XPO
concentrator to its AC or
2
DC power adapter, or to its supplemental
battery pack.
2.
If no other power source is available,
change to a reserve oxygen supply.
1148112-F~12