Adjusting The Pulse Flow Setting - Invacare XPO100 User Manual

Portable oxygen concentrator
Hide thumbs Also See for XPO100:
Table of Contents

Advertisement

Available languages

Available languages

Invacare® XPO₂™
Altitude:
Tubing and Cannula:
Environment:
Time of Operation:

5.4 Adjusting the Pulse Flow Setting

FLOW SETTING
1
2
3
4
5
24
Up to 10,000 ft (3046 m) above
sea level.
4 ft (1.2 m) cannula with a
maximum 25 ft (7.6 m) of crush
resistant tubing (DO NOT
pinch).
Smoke, pollutant and fume free.
No confined spaces (example:
no closets).
Up to 24 hours per day when
connected to AC or DC power
adapter.
AVERAGE PULSE
VOLUME (AT 20
BREATHS PER MINUTE)
16 ml
24 ml
31 ml
37 ml
42 ml
1.
Turn On the XPO
concentrator by pressing and holding the
2
power On/Off button A for approximately one second. Each
time the XPO
concentrator is turned On, a brief alarm will
2
sound. The XPO
concentrator pulse flow setting at power-up
2
will be the same flow setting selected the last time the unit was
turned Off.
2.
The XPO
concentrator has five pulse flow settings. Press the
2
flow selection (+ or -) button B until your prescribed flow
setting (1 to 5) is illuminated by a blue indicator light C.
WARNING!
Risk of Injury
It is very important to select the prescribed level of
oxygen flow.
– DO NOT increase or decrease the flow unless a
change has been prescribed by your physician or
therapist.
1148112-F~12

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Xpo100b

Table of Contents