Annexe A Neox Standard - Settings And Restrictions - Physipro NeoX Owner's Manual

Hide thumbs Also See for NeoX:
Table of Contents

Advertisement

ENGLISH
Appendix A : NEOX Standard — Settings and Restrictions
This document appendix describe the range of adjustment available and limits on the NEOX
tilt in space. It covers the following sections:
• Adjustment sequence
• Seat depth adjustment
• Center of mass adjustment
• Gas cylinder adjustment
• Seat to floor height adjustment / limitations
• Tilt / recline stability limitation
• Overall width and tilt limitation
Adjustment sequence
In order to properly adjust the NEOX Tilt-in-Space, the following sequence should be applied:
• Adjust the seat width according to order form
• Adjust the seat depth according to order form and reference table
• Adjust the center of mass according to reference table
• Adjust gas cylinder according to reference table
• Adjust seat to floor height according to order form
• If required, limit the tilt/recline according to reference table
Seat depth adjustment
Adjust the seat depth according to Table 1. This table consider that the interface between
the back and the patient is as define on the Figure 1. A thicker cushion or recessed back will
change the adjustment.
Back / patient
Backrest/patient
Interface
interface
Figure 1: Seat depth adjustment--standard model
50
NEOX™ Tilt-in-Space

Advertisement

Table of Contents
loading

Table of Contents