XSYNC Connector (Figure 2-11)

Removal

1. Remove the rear panel and XSYNC assembly.
2. Using a small flat scredriver, remove the screw securing the XSYNC cable to the XSYNC
assembly.
Note: XSYNC assembly consists of:
XSYNC assembly (connector)
XSYNC cable and screw
Nylon shoulder washer
3. Using the XSYNC nut driver, remove the nut and lug cable asssembly that secures the XSYNC
assembly to the rear panel.
Note: Nut and lug assembly consists of:
nut
Inner shoulder washer
XSYNC lug cable
4. Lift off the XSYNC assembly and nut and lug cable assembly from the rear panel.
Screw
XSYNC Cable
Rear Panel
Nut
XSYNC
Lug Cable
Figure 2-11. Replacing XSYNC assembly
35
Nylon Shoulder
Washer
Assembly
Inner Shoulder
Washer
Parts Replacement
XSYNC