Polaroid Digital Microscope Camera Repair Manual page 50

Hide thumbs Also See for Digital Microscope Camera:
Table of Contents

Advertisement

DMC Repair Manual
Flair & Bloom
• Click on the Flair & Bloom button. A window appears on the Pretest VI panel indicating
that the Flair and Bloom test is running. It also prompts you place a piece of white
paper approximately 1 foot in front of the camera to check for dark spots. this test
also checks foil for flair bloom.
• Once white paper is placed in front of the camera, click on the OK button. A white
screen appears. Check for dark spots.
Shutter Light Leak
• Turn on Lamp.
• Click on the Shutter Light leak button. The Shutter Alignment window appears.
Also, a dark image of the target appears with a bright red LED in its middle.
• Turn off Lamp.