DMC Repair Manual
13. Set the light box to 50 Cd/Ft
Note:
The calibration table affixed to the light box indicates the equivalent diode current
for the particular light box setting. Adjust the light until the Keithly 485
AutoRanging Picoammeter reads the specified diode current.
14. Once the light box is set to 50 Cd/Ft
This action starts the selected LabView vi calibration program.
15. Upon completion of the ambient portion of the calibration test program, a message
appears prompting you to move the camera to the focus test station (Diorama Setup of
Macbeth Chart and Sparkly Foil).
Make sure that the power cord is not disconnected when
moving the camera from the ambient test station to the
focus test station.
If power is disconnected, you must restart the calibration
test from the beginning.
2
.
2
(candles per foot), click on the Start button.
Caution
58
Testing & Adjustment
Need help?
Do you have a question about the Digital Microscope Camera and is the answer not in the manual?
Questions and answers