Setting
Details
items
F17
Not used
F18
Beep sound
F19
Not used
External
F20
input/output unit
(option) protocol
Transmission
F21
speed
(Mini-DIN)
Transmission
F22
speed
(D-Sub)
Stop bit
F23
(Mini-DIN)
Stop bit
F24
(D-Sub)
Blood pressure
F25
result output
F26
Date format
F27
Time format
F28
Not used
F29
ICT printing
To reset all settings to factory settings, hold the START/STOP button for 5 seconds when
any of the "FXX" numbers are displayed.
The default setting depends on the product model.
Diastolic
Default
display
section
/
21
Beep sound
on/off
No connection
Mini-DIN:
blood pressure result output (STD/RI/RB/BP/RA)
D-Sub:
blood pressure result output (STD/RI/RB/BP/RA)
Mini-DIN: connect to the A&D weight scale
D-Sub:
blood pressure result output (STD/RI/RB/BP/RA)
Mini-DIN:
blood pressure result output (STD/RI/RB/BP/RA)
D-Sub: ID reader
Mini-DIN:
blood pressure result output (STD/RI/RB/BP/RA)
D-Sub: Ux compatibility
Mini-DIN:
blood pressure result output (STD/RI/RB/BP/RA)
D-Sub: RVX compatibility
Mini-DIN:
blood pressure result output (STD/RI/RB/BP/RA)
D-Sub: connect to the A&D weight scale
1200 bps
2400 bps
4800 bps
9600 bps
1200 bps
2400 bps
4800 bps
9600 bps
Stop bit: 1
Stop bit: 2
Stop bit: 1
Stop bit: 2
RB (no ID, immediately after measurement) + STD
RI (with ID, immediately after measurement)+ STD
BP (with ID, immediately after measurement)only
STD (command response) only
RA (with ID, immediately after measurement)
DD month., YYYY
(TM-2657-EX)
month. DD, YYYY
(TM-2657-EG)
24 hour
(TM-2657-EX)
12 hour (AM/PM)
(TM-2657-EG)
ICT printing OFF
Bar code printing (CODE39)
QR code printing
Function