In order to use the TM-2657P Automatic Blood Pressure Monitor safely and correctly, carefully read
the following precautions before using the monitor. The following content summarizes general matters
regarding the safety of patients and operators, in addition to safe handling of the monitor.
1. When installing and storing the monitor.
Keep the monitor away from areas where flammable anesthetics or flammable gases
are present, high-pressure oxygen chambers, and oxygen tents. Using the monitor in
these areas may cause an explosion.
Consider the following when using and storing the monitor. If the monitor is stored in an
environment beyond the specified temperature or humidity, it may not perform to its
capabilities.
Avoid locations where the monitor may be splashed by water.
Avoid locations with high temperature, high humidity, direct sunlight, dust, salt and
sulfur in the air.
Avoid locations where the monitor may be tilted, vibrated, or impacted (including
during transportation).
Avoid locations where chemicals are stored or gas occurs.
Installation site: A location with a temperature between +10
humidity between 15% RH and 85% RH (no condensation).
Storage site: A location with a temperature between -20
between 10% RH and 95% RH.
A location with an electrical outlet that can supply sufficient power (frequency,
voltage, current) for the monitor.
Please be aware that the rubber feet may discolor the top of the stand.
PRECAUTIONS FOR USE
Danger
Caution
Note
ii
and +40
and a
°C
°C
and +60
and a humidity
°C
°C