SERVICE MANUAL
8.4.5
Checking the cables
Requirements:
● Side cutting pliers
● Cable ties
●
Remove the covers, disconnect the battery plug connections and pull the batteries out of the
battery compartment, as described in chapter 8.4.1.
●
Check the fuse cable (1), the battery cable (2) and the battery connections (3) for visible
damage and pinches.
●
Replace damaged cables, as described in chapter 8.4.2.
Note
The positions of the plugs are described in chapter 4.2.
56
®
Invacare
- FDX