Invacare Bora XTR Series Service Manual page 30

Hide thumbs Also See for Bora XTR Series:
Table of Contents

Advertisement

SERVICE MANUAL
Problem
Batteries
not being
charged
Mobility
device
runs too
slowly
Electrical
adjustment
motor
does not
react
30
Other
Possible
Symptoms
Cause
None
Batteries
defective
LEDs blinking
Charging unit
on charging unit
defective
None
Remote
defective
Batteries
defective
Remote shows
Lighting /
a blinking "E",
actuator
status diode on
module
lighting/actuator
defective
module does
not go out even
if remote is
switched off or
disconnected.
None
Cable
disconnected
or damaged
Electrical
adjusting
motor
defective
Remote
defective
®
®
Invacare
- Bora
/Spectra
Solution
Documentation
Replace carbon
See chapter 7.2.4.
brushes
Replace batteries
See chapter 7.7.
Replace charging
See charging unit
unit
operating manual
Replace remote
See chapter 7.10.
Replace batteries
See chapter 7.7.
Replace lighting /
See chapter 7.2.9.
actuator module
Safeguard cable
See chapter 7.9.
connection,
replace cable if
necessary
Check adjusting
See chapter 7.17.
motor
Replace remote
See chapter 7.10.
®
XTR Series

Advertisement

Table of Contents
loading

This manual is also suitable for:

Spectra xtr series

Table of Contents