Pull the power cord through the opening on the power cord receptacle.
Using same distance located on old power cord, slide the strain relief
onto the new power cord with the lip of the strain relief facing the ends
of the power cord leads.
Insert the new power cord through the opening on the receptacle by
squeezing the strain relief together with pliers.
Position the strain relief upward, and push it into the rear of the power
cord opening on the receptacle.
Reconnect the power cord leads, close the plastic twist clamps, and
reattach the tie-wraps.
Reconnect the power cord receptacle to the base of the unit with the
two Phillips-head screws.
Reconnect the side and back panels.