Feed Or Waste Valve Reassembly; Solenoid Valve Coil Replacement - AirSep NewLife Elite Service Manual

Oxygen concentrator
Hide thumbs Also See for NewLife Elite:
Table of Contents

Advertisement

NewLife
Elite Service Manual
®

5.4.2 Feed or Waste Valve Reassembly

5.4.3 Solenoid Valve Coil Replacement

5-5
Loosen and remove the solenoid base with a one-inch deep
5
well socket.
Install the rebuild kit, which contains all parts of the sole noid
6
assembly except the solenoid coil.
To reassemble the feed or waste valves, follow the feed or waste valve
rebuilding procedure in reverse order, and test for leaks.
If a solenoid valve coil does not operate, its corresponding green light
on the circuit board does not illuminate.
Set the ON/OFF switch to the OFF position, and unplug the
1
power cord.
2
Remove the side and back panels.
Remove the retaining cap with a slotted-head screwdriver.
3
Disconnect the solenoid leads, and lift off the solenoid coil.
4
Replace with the new coil.
5
Reinstall the retaining cap back on top of the coil, and reconnect
6
the solenoid leads.
7
Reconnect the back and side panels.