5.5

Sieve Bed Replacement

5.5.1
AirSep
Corporation
®
WARNING
Do not refill the snap-ring-style sieve beds. Special
tooling is required to remove and re-install the spring-
loaded top cap and snap-ring.
CAUTION
Do not expose molecular sieve (contents of bed) to air
for an extended period of time. Pro longed exposure
of molecular sieve to the moisture in room air results
in contamination and permanent damage to the sieve
material. Keep all openings to the sieve beds sealed.
NOTE
If replacement is necessary, you must replace both sieve
beds at the same time.

Sieve Bed Removal

Set the unit's ON/OFF switch to the OFF position, and unplug
1
the power cord.
2
Remove the side and back panels.
Cut the tie-wraps at the brass "T" fitting, and discon nect the
3
green 1/4-inch product tubes.
Cut the tie-wraps at the top of the sieve beds, and disconnect
4
the green 1/4-inch equalization tubes.
Remove the 9/16-inch compression fitting on the bottom of
5
each sieve bed.
6
Cut the tie-wraps, and remove the sieve beds. Proper removal
of the sieve beds includes the product tubes.
7
Plug the opening on the top and bottom of the sieve beds.
NewLife
Elite Service Manual
®
5-6