5.11.2 ON/OFF Power Switch Installation

5.12 Buzzer Replacement

1
2
3
4
5

5.13 Hour Meter Replacement

1
2
3
4
5
6
7
8
5-15
Follow the removal procedure for the ON/OFF power switch in reverse
order to install a new power switch. Make sure to reinstall the switch
properly by having the "O" on the switch located on the bottom when
finished.
Set the unit's ON/OFF switch to the OFF position, and unplug the
power cord.
Remove the left side panel.
Unscrew the two Phillips-head screws that hold the buzzer to the
superstructure, and solder to disconnect the buzzer leads.
Solder the leads to the new buzzer, and screw the buzzer to the
superstructure.
Reconnect the side panel.
Set the unit's ON/OFF switch to the OFF position, and unplug the
power cord.
Remove the left side panel.
Disconnect the hour meter leads.
Push hour meter mounting tabs to remove the hour meter through the
control panel.
Install the new hour meter into the control panel, making sure that the
"hourglass" icon displays on the bottom.
Reconnect the hour meter leads.
Reconnect the side panel.
Plug unit in and set the unit's ON/OFF switch to ON. Hourglass icon
should flash showing it is functioning correctly.