5.14 Flowmeter Replacement

5.14.1 Flowmeter Removal

5.14.2 Flowmeter Installation

5.15 Power Cord Replacement

1
2
3
4
5
6
7
8
9
AirSep
Corporation
®
Set the unit's ON/OFF switch to the OFF position, and unplug
1
the power cord.
Remove the left side panel.
2
Cut the tie-wraps, and remove the 3/8-inch green oxygen
3
tubing from the flowmeter fittings.
Unscrew the flowmeter nuts with a wrench.
4
Remove the flowmeter through the control panel.
5
To install a new flowmeter, follow the flowmeter removal proce dure
in reverse order. Then perform a leak test on the connections.
Set the unit's ON/OFF switch to the OFF position, and unplug the
power cord.
Remove the side and back panels.
Open the plastic twist clamps.
Cut the tie-wraps (located near the power cord push-on connectors).
Disconnect the white and black power cord leads.
Lay the unit face down, and unscrew the two Phillips-head screws that
secure the power cord receptacle to the base of the unit.
Use a slotted-head screwdriver to break the receptacle seal.
Pull the receptacle forward until the back of the strain relief is visible.
Press both ends of the strain relief together with needle-nose pliers to
remove and save it for reuse.
NewLife
Elite Service Manual
®
5-16