5.10 Circuit Breaker Replacement

5.10.1 Circuit Breaker Removal

5.10.2 Circuit Breaker Installation

5.11 ON/OFF Power Switch Replacement

5.11.1 ON/OFF Power Switch Removal

AirSep
Corporation
®
Install the new pressure switch, connect the leads, and leak test.
Reconnect the side panel.
Set the unit's ON/OFF switch to the OFF position, and unplug
1
the power cord.
Remove the left side panel.
2
Disconnect the circuit breaker leads.
3
Unscrew the circuit breaker while you apply pressure to the
4
circuit breaker retaining ring.
Follow the removal procedure for the circuit breaker in reverse order
to install the new circuit breaker.
Set the unit's ON/OFF switch to the OFF position, and unplug
1
the power cord.
Remove both side panels, the lower front panel, and the back
2
cover.
3
Unscrew the four Phillips-head screws that hold the control
panel to the superstructure.
Cut off the two white tie-wraps from the terminal strip leads.
4
Disconnect the ON/OFF switch leads from the back of the
5
switch.
Push on the back of the power switch, while holding in its
6
retaining tabs, and remove the switch through the front panel.
NewLife
Elite Service Manual
®
5-14