AirSep SeQual Regalia Instruction Manual

Psa oxygen concentrator system

Advertisement

Quick Links

SeQual
Regalia
®
PSA Oxygen Concentrator System
INSTRUCTION MANUAL

Advertisement

Table of Contents
loading

Summary of Contents for AirSep SeQual Regalia

  • Page 1 SeQual Regalia ® PSA Oxygen Concentrator System INSTRUCTION MANUAL...
  • Page 2: Table Of Contents

    Tables of Contents 1.0 Introduction GENERAL CAUTIONS, WARNINGS AND HAZARDS 2.0 Operation GENERAL RECOMMENDED OPERATING ENVIRONMENT GUIDELINES 11 THE PROPER LOCATION BEFORE OPERATING YOUR OXYGEN CONCENTRATOR OPERATING YOUR OXYGEN CONCENTRATOR 3.0 Installation GENERAL 4.0 Power Management Feature IN THE EVENT OF A POWER INTERRUPTION: 5.0 Operating Conditions CONTROL PANEL DISPLAY DISPLAY OF OPERATING CONDITIONS...
  • Page 3 TECHNICAL SUPPORT...
  • Page 5 List of Figures and Tables Table 2.1: Regalia Operating Environment Guidelines Table 5.2: Display of Operating Conditions Table 6.3: Concentrator is NOT Running Table 6.4: Concentrator is Running Table 7.5: Regalia Specifications...
  • Page 6 Symbols can also permit easier comprehension of a concept within a restricted space. The following table is a list of symbols and definitions that may be used with AirSep oxygen equipment. These symbols are referenced from the appropriate International Electro-technical Commission (IEC) standards:...
  • Page 7 Class II Equipment - Equipment in which protection against electrical shock does not rely on Amperes basic insulation only, but in which additional safety precautions are provided. Note – Provides information important enough to emphasize Alternating Current or repeat SeQual Regalia Instruction Manual...
  • Page 9: Introduction

    ■ Do NOT cover unit with a blanket, towel, quilt, etc. ■ Do NOT place objects (pillow, blanket) under unit. Maintain at least ½ inch clearance under unit. SeQual Regalia Instruction Manual...
  • Page 10 Patents, Trademarks and Copyright The Oxygen Concentrator System is protected by the following US Patents: ■ 4,925,464 ■ 5,112,367 ■ 5,114,441 ■ 5,268,021 ■ 5,366,541 ■ 5,593,478 ■ 5,730,778 ■ Re. 35,099 ■ Other patents pending SeQual Regalia Instruction Manual...
  • Page 11: Operation

    From 0.5 LPM to 10.0 LPM Table 2.1: Regalia Operating Environment Guidelines For rated performance at 10 LPM do not operate the oxygen concentrator above the indicated boundaries. Operating at flow rates less than 10 LPM will tolerate higher temperature/humidity conditions. SeQual Regalia Instruction Manual...
  • Page 12: The Proper Location

    The red and/or yellow light may stay on for up to ten (10) minutes after start-up, OM models only. This is a normal condition during the initial warm-up period. SeQual Regalia Instruction Manual...
  • Page 13: Operating Your Oxygen Concentrator

    OPERATING YOUR OXYGEN CONCENTRATOR This manual serves as a supplemental reference to help you operate and maintain the unit. If you have any further questions or problems call your AirSep dealer or authorized service representative. General Precautions – Retain the product’s safety and operating instructions for future reference.
  • Page 15: Installation

    If the unit is equipped with an oxygen concentration monitor (OM models only), the red, yellow and green light will intermittently illuminate while the unit is warming up. For details on warm-up refer to the Display of Operating Conditions table. SeQual Regalia Instruction Manual...
  • Page 16 5. Flow Rate Adjustment Adjust the flow rate to desired setting. Press the Plus “+” button to increase or the Minus “+” button to decrease the flow rate until the proper level is displayed. SeQual Regalia Instruction Manual...
  • Page 17: Power Management Feature

    If the unit was not turned off the Regalia will automatically resume operation once power is restored. If the unit was turned off during the power interruption, the user must press the ON/OFF button again, after the power is restored, to turn the unit on. SeQual Regalia Instruction Manual...
  • Page 19: Operating Conditions

    Regalia is running and oxygen is being produced. If the concentration level falls below preset levels, an indicator light will activate. Operating conditions are described in the table on the following page. SeQual Regalia Instruction Manual...
  • Page 20: Display Of Operating Conditions

    If abnormal conditions are detected and cannot be remedied, refer to the Troubleshooting section. During System Malfunction condition, the Green and Yellow lights may be On or Off. Table 5.2: Display of Operating Conditions NON-USAGE When not using the oxygen concentrator, press the On/Off button to turn it off. SeQual Regalia Instruction Manual...
  • Page 21: Maintenance

    Do Not operate the oxygen concentrator for more than 30 minutes without a filter installed. OXYGEN DELIVERY ACCESSORIES Follow the instructions supplied by the manufacturer. SeQual Regalia Instruction Manual...
  • Page 22: Troubleshooting Charts

    Remove any obstruction(s) Audible Alarm Unit overheating from air openings and refer Blinking Red Light to “The Proper Location”. Flow Indicator Blank Back Light On Compressor malfunction Service required. Table 6.3: Concentrator is NOT Running SeQual Regalia Instruction Manual...
  • Page 23 Flow Rate not visible on Flow LCD malfunction Service required. Indicator Display Audible Alarm Remove power cord from Red Light On wall outlet. System malfunction Flow Indicator Blank Service required. Back Light On Table 6.4: Concentrator is Running SeQual Regalia Instruction Manual...
  • Page 25: Technical Data

    SCF (Standard cubic foot) gas measured at 1 atmosphere and 70°F / Nm (Normal cubic meter) gas measured at 1 atmosphere and 0°C / LPM (Liters per minute) gas measured at 1 atmosphere and 21°C Table 7.5: Regalia Specifications SeQual Regalia Instruction Manual...
  • Page 26: Disposal Of Equipment And Accessories

    TECHNICAL SUPPORT For assistance in troubleshooting or repairing the unit, or to order replacement parts, contact the AirSep Commercial Products Service Department by telephone. In the USA or Canada, call 1-800-320-0303. Outside the USA or Canada, call (716) 691- 0202. Send fax inquiries anytime to (716) 564-2717. Address written inquiries to:...
  • Page 28 © 2019 AirSep Corporation. All Rights Reserved. AirSep reserves the right to discontinue its products, or change the prices, materials, equipment, quality, descriptions, specifications and/or processes to its products at any time without prior notice and with no further obligation or consequence.

Table of Contents