5.8.3 Back Pressure Correction at 5 LPM

5-11
NOTE
It may be desirable to provide one addi tional
liter of flow if you use the lockout feature.
1
Plug in the unit.
Set the unit's ON/OFF switch to the ON position, and allow
2
the unit to run at least three minutes to build up pressure.
If used, connect the humidifier bottle, oxygen tubing, and
3
cannula to the oxygen outlet.
Remove the right side panel.
4
Turn the flowmeter adjustment knob counterclockwise un til
5
it stops (wide open).
Pull outward on the regulator knob to unlock it.
6
Turn the regulator knob counterclockwise to decrease the
7
setting until the flowmeter registers the maximum flow
desired.
Push in the regulator knob to lock the setting.
8
9
Reconnect the side panel.
In some cases, additional lengths of oxygen tubing with a humidi fier
bottle can increase back pressure and limit oxygen flow below 5 lpm.
To achieve 5 lpm, use the following procedure:
Plug in the unit.
1
NOTE
When this lockout feature is no longer
required, follow the normal regulator
setting proce dure, as described in
Section 5.8.1.
AirSep
Corporation
®