Omron HBP-1100 Instruction Manual page 21

Professional blood pressure monitor
Hide thumbs Also See for HBP-1100:
Table of Contents

Advertisement

Caution
• The accuracy of a ashing measurement value that is outside the
measurement range is not guaranteed. Always check the patient's
condition before deciding what steps to take.
NOTE:
• If in ation is insuf cient, in ation may restart automatically while
measurement is in progress.
Irregular pulse wave detection function
If the pulse wave interval becomes irregular during measurement, the
irregular pulse wave detection icon will light to notify you.
Motion stop function
If body movement is detected during measurement, de ation stops for 5
seconds and "-" "|" blink below the value.
After 5 seconds, measurement resumes, and an attempt is made to
complete measurement in one cycle.
NOTE:
• When the motion stop function has activated, the irregular pulse wave
icon appears in the measurement result.
19